Filters should be used during the administration of Parenteral Nutrition for patients in all care settings DSI/2026/011
Safety Classified by Officially
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome.
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome. Healthcare professionals are advised to use integrated or add-on filters during the administration of parenteral nutrition for patients in all care settings. All parenteral nutrition solutions should be administered via a filter with a pore size no larger than 1.2μm. Where available, it is good practice to administer aqueous/lipid-free parenteral nutrition solutions with a 0.2μm filter.
Advice for Healthcare Professionals to Provide to Patients:
This is an extract. The publication continues at the source.
Read the original at the source: https://www.gov.uk/drug-device-alerts/filters-should-be-used-during-the-administration-of-parenteral-nutrition-for-patients-in-all-care-settings-dsi-slash-2026-slash-011
Officially imported this from Medicines and Healthcare products Regulatory Agency’s own source and shows an extract. If you work there, claiming the profile and verifying the domain lets you choose to show the full text here.
Provenance
- Organization
- Medicines and Healthcare products Regulatory Agency — imported from official source
- Official source
- https://www.gov.uk/government/organisations/medicines-and-healthcare-products-regulatory-agency.atom ATOM
- Imported
- September 16, 2026 05:30
- Versions
- 1 recorded
- Identity
https://www.gov.uk/drug-device-alerts/filters-should-be-used-during-the-administration-...