National Patient Safety Alert: Patient hoists and slings (all types): risk of death and serious harm from falls (NatPSA/2026/005/MHRA)

Imported from official source

Alert

Safety Classified by Officially

Fatal and serious harm continues to occur when patients fall from hoists or slings during transfers and repositioning. A review of recent incidents with a fatal outcome and wider surveillance data shows an average of two deaths per year since 2015.

This alert is for action by: All those responsible for the use, purchase and/or maintenance of patient hoists, working within all acute and community healthcare organisations, care homes, nursing homes and care services supporting people in their own homes, and their equipment providers.

This is a safety critical and complex National Patient Safety Alert. Implementation should be coordinated by an executive leader
(or equivalent senior accountable person, such as a registered manager or responsible induvial, in organisations without executive boards).

Fatal and serious harm continues to occur when patients fall from hoists or slings during transfers and repositioning. A review of recent incidents with a fatal outcome and wider surveillance data shows an average of two deaths per year since 2015. This is not confined to any manufacturer, model, or care setting. The most common causes of fatal and serious patient falls from hoists include:

  • Detachment at critical load-bearing interfaces, including spreader bar hooks, clips, carabiners, fasteners, or other parts that are missing, worn, damaged or incorrectly assembled.
  • Incompatible hoist and sling combinations, including use of third-party slings not validated for the specific hoist.
  • Failure to identify damaged or incorrectly seated sling loops, or incorrect attachment to spreader bars during fitting and use.
  • Failure to conduct effective pre-use checks that would identify unsafe equipment.
  • Inadequate or overdue maintenance, servicing and examination under the Lifting Operations and Lifting Equipment Regulations 1998 (LOLER). This included continued use of a hoist or component parts beyond their indicated service life without risk assessment.
  • This is an extract. The publication continues at the source.

    Read the original at the source: https://www.gov.uk/drug-device-alerts/national-patient-safety-alert-patient-hoists-and-slings-all-types-risk-of-death-and-serious-harm-from-falls-natpsa-slash-2026-slash-005-slash-mhra

    Officially imported this from Medicines and Healthcare products Regulatory Agency’s own source and shows an extract. If you work there, claiming the profile and verifying the domain lets you choose to show the full text here.

    Provenance

    Organization
    Medicines and Healthcare products Regulatory Agency — imported from official source
    Official source
    https://www.gov.uk/government/organisations/medicines-and-healthcare-products-regulatory-agency.atom ATOM
    Imported
    September 16, 2026 11:30
    Versions
    1 recorded
    Identity
    https://www.gov.uk/drug-device-alerts/national-patient-safety-alert-patient-hoists-and-...

    Officially records where a publication came from, not whether it is true. Imported records are reproduced from an organization's own official source.